Темир йўл транспорти: долзарб масалалар ва инновациялар 2-jild (2024) · 77-83-betlar
ERRORS OF A NEW INDUCTION VIBRATION TRANSDUCER: ERRORS OF A NEW INDUCTION VIBRATION TRANSDUCER
Файзуллаев, Ж.С., Шоимкулов, A.A.
Annotatsiya
The paper shows that to reduce the amplitude dynamic error of a new induction vibration transducer, the input of which is input harmonic measured value (vibration oscillations) it is necessary to increase the natural frequency of the measuring device and select the optimal value of the degree of relative quiescence, and to reduce the phase dynamic error it is necessary to select the minimum, close to zero, value of the degree of relative quiescence. The process of selecting the optimal parameters of the new induction vibration transducer in order to reduce its amplitude and phase dynamic error has shown that it is practically impossible to ensure their simultaneous reduction, explained by the fact that the amplitude frequency and phase-frequency characteristics are mutually related.
индукционный преобразователь вибраций, основная погрешность, дополнительная погрешность, амплитудная относительная погрешность, фазовая относительная погрешность, динамическая характеристика, амплитудно-частотная характеристика, фаза-частотная характеристика, собственная частота, степень успокоения.induction vibration transducer, basic error, additional error, amplitude relative error, phase relative error, dynamic characteristic, amplitude-frequency characteristic, phase-frequency characteristic, natural frequency, degree of calming